Centrifuge 5804/5804 R/5810/5810 R — Operating manual. Emergency release. If the centrifuge lid cannot be opened, you can activate the emergency release manually. You need the standard rotor key supplied with the centrifuge. Disconnect the power plug. Remove the plastic cover for the emergency release. This is located in the center. 9 Centrifuge 5804/5804 R/5810/5810 R — Operating manual 1 User instructions 1 User instructions 1.1 Using this manual Read this operating manual completely before using the device for the first time.Please also note the operating instructions for the accessories, if applicable.

DESCRIPTION
Unparalleled versatility.

Featuring powerful, maintenance-free motors, these models accommodate a variety of rotors to meet your application needs. They are three centrifuges in one: a high-capacity, general-purpose centrifuge for cell harvesting; a high-speed centrifuge for separating cell lysates; and a microcentrifuge for DNA precipitations.
Model 5810 and refrigerated Model 5810 R* offer maximum application versatility and popular tube capacity, accommodating up to 4 x 400 ml tubes or 16 microplates. These compact, benchtop centrifuges achieve remarkable throughput of up to 48 x 15 ml or 20 x 50 ml conical (Falcon®) tubes

FORUMSView All (11)
CentrifugeCentrifuge

Eppendorf Centrifuge 5810 R Service Manual

Ask a New Question

3

Replies
-doowon
6 months ago
6 months agoError code 31
Hello, have some issue with error code 31. when it started it door latch is hunting and ends up with error code 31.I replaced two sensors on the door latch but did not help. changed PCB board with brand new same thing happens. Anyone have any ideas?Reply
-paul.blackaby_3666
a year ago
a year agoService Manual
Anyone able to supply a service manual for a 5810R or the method of changing the lid strut. Thankyou in advanceReply

2

Replies
-TechST
2 years ago
2 years agoProcedure
Hi to all, i need replace the main board of this device and from Eppendorf says this procedure can do it only the oficial service because when you replace this PCB is necesary adjust the imbalance sensor. If in this case the information it's true:How can i do that procedure? Is posible to explain me?Thanks in advance...Reply

FEATURES

  • Interchangeable swingbucket and high-speed fixedangle rotors
  • Spin up to 14,000 rpm, or 20,800 x g
  • Low-profile 11.2 in/28 cm access height for easy sample loading
  • Automatic rotor recognition senses rotor type to set maximum allowable speed—ensures safety
  • Fast, simple programming of time, temperature (refrigerated models), speed/ g-force and radius value
  • 34 user-defined programs
  • Adjustable-speed Short Spin button
  • Ten acceleration and braking rates
  • Simple operation
  • Speed, g-force and radius correction values can be entered
  • Values can be changed during centrifugation
  • Temperature range from –9 °C to 40°C, with Fast Temp function for rapid cooling of the centrifuge chamber
  • Standby cooling maintains temperature when not in use
  • Speed can be set from 200 rpm to maximum speed in increments of 10 rpm
  • Short Spin button
  • Selectable program memory for up to 34 individual user programs
  • Ten acceleration and deceleration rates for sensitive sample material
SERIES MODELS
Height13 in
Length 24 in
Weight150 lbs
Width21 in
Additional SpecificationsMaximum power requirement: 900 W

Eppendorf 5810 Centrifuge

Height13 in
Length 24 in
Weight218 lbs
Width28 in
Additional SpecificationsMaximum power requirement: 1,650 W